The FHT 59 E continuously monitors beta emitting noble gases
in workplace or effluent air. The sampled air is drawn through a measuring vessel with the
emitted radiation being detected by two large area proportional counter tubes.
application in workplace or stack emissions monitoring systems
calibration factors for Kr-85 and Xe-133 available
no gas supply required by virtue of sealed proportional counter tubes
maintenance-free long term operation
direct read-out of measured values in physical units of Bq/m3
easy systems integration via RS232 serial interface
LAN coupling optionally available
Features
measuring range 0.5/1.0× 103Bq/m³ - 108Bq/m³
for Kr-85 and Xe-133, respectively
detector assembly comprising two large area proportional counter tubes in sandwich
geometry
dynamic compensation of varying levels of ambient doserate by subtraction of crosswise
anticoincident detector countrates
minimised self absorption by virtue of a flat detection cuvette
aerosol filter
measuring electronics FHT8000 operable under windows
menu driven calibration check procedure with complete track record of detector data