| The FHT 63 D
continuously monitors tritium in workplace or effluent air. The counting gas and the
sample air are blended and measured inside a triple proportional counter tube:
- application in stack emissions monitoring systems
- dynamic compensation of interfering noble gas isotopes
- low-cost methane gas supply, not necessitating counting gas in high level spectroscopy
quality
- long term maintenance free operation
- direct read-out of measured values in physical units of Bq/m³
- easy systems integration via RS232 serial interface
- LAN coupling optionally available
Features
- measuring range for tritium 700 108 Bq/m³
- gas flow triple proportional counter tube
- aerosol filter
- measuring electronics FHT8000 operable under windows
- menu driven calibration check procedure with complete track record of detector data
- low noise maintenance free membrane pump; sample air flow 11.8 l/h
|
|